Ambient and process temperature limits for CMFS007, CMFS025–CMFS150
140 (60)
–40 (–40)
113
(45)
–148 (–100)
–58 (–50) 400 (204)
140 (60)
A
B
B
Tamb
Tproc
T
amb
= Ambient temperature °F (°C)
T
proc
= Process temperature °F (°C)
A = All available electronic options
B= Remote mount electronics only
Ambient and process temperature limits for CMF***M/L/H/P (excludes special order
cryogenic modifications) and CMFS010-015
140 (60)
–40 (–40)
113
(45)
–148 (–100)
–400
(–240)
400
(204)
140
(60)
T amb
T proc
A
B
B
C
-148
(-100)
T
amb
= Ambient temperature °F (°C)
T
proc
= Process temperature °F (°C)
A = All available electronic options
B = Remote mount electronics only
C = Recommend special order cryogenic sensor options when operating at a process
temperature below -148 °F (-100 °C)
Planning Installation Manual
October 2021 20002158
10 Micro Motion ELITE

Other manuals for Emerson Micro Motion ELITE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion ELITE and is the answer not in the manual?

Emerson Micro Motion ELITE Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion ELITE
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals