Attribute Access Name Type Trigger Byte
WriteAttribute
Description
19h Get Smart Meter Verification
Progress
U16 13h % complete
1Ah Get API Table Type U16 14h See Table B-18
1Bh Get Reference temperature Float 15h
1Ch Get Thermal expansion coefficient
(TEC)
Float 15h
1Dh Get Subnet communication status
(RS-485 connection to
transmitter)
U16 No trigger byte 0: Communications failure
Any other value:
Communications good
B.3.12 Output parameters (explicit data) for petroleum
measurement configuration
Class B1h, Instance 01h
Attribute Access Name Type Trigger byte
write attribute
Description
01h Get/Set Standard or special mass flow
rate unit
U16 See Table B-7
02h Get/Set Density unit U16 See Table B-13
03h Get/Set Temperature unit U16 See Table B-14
04h Get/Set Standard or special volume
flow rate unit
U16 See Table B-10
05h Get/Set Pressure unit U16 See Table B-15
06h Get/Set Zero time U16 Seconds
07h Get/Set Zero value Float
08h Get/Set Output state during Smart
Meter Verification
U16 0 = Last measured value
1 = Fault
09h Get/Set Smart Meter Verification Index U16 16h Smart Meter Verification test
record.
0 = most recent
...
19 = oldest
0Ah API Table Type U16 See Table B-18
0Bh Reference temperature Float
User Guide Device profile
MMI-20019808 January 2019
User Guide 55