Instruction Sheet
748336-C
April 1997
Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 I/O Module
4 of 4 Addendum
4. The screen will return to Module Bind-
ing, and the I/O Module(s) selected will
appear as Proposed New Bind.
5. If the I/O Module in Proposed New Bind
is correct, press the <BIND> softkey.
(The system resets itself automatically
when the Bind softkey is pressed.)
6. If the I/O Module in Proposed New Bind
is not correct, return to Select Modules
and select the correct I/O Module(s).
7. To bind additional Analyzer and I/O Mod-
ules, return to the Run Mode screen,
press <NEXT> softkey, select another
Analyzer Module and repeat steps 2, 3,
and 4.
As I/O Modules are bound, they become un-
available for binding, and are automatically
removed from the Select Modules screen. In
a system consisting of one Analyzer Module,
I/O Modules are automatically bound upon ini-
tialization of the system. Therefore no I/O
Modules would appear in the Select Module
screen.
A "miss-bound" module can lead to confusion
and misdiagnosis of system problems. Verify
all bindings by selecting Module Binding,
View Bindings and using the <NEXT> softkey
to toggle through Analyzer-I/O Module binding
combinations.
At this point, all NGA 2000 system compo-
nents are bound and fully coordinated.