41
Safety Manual
00809-0200-5100, Rev GB
Installation and Configuration
February 2017
Installation and Configuration
4. Perform a final verification of the following parameters:
Check that SIL High Alarm Limits are within approved limits, see “Setting the SIL High
Alarm Limit” on page 27.
If SIL Low Alarm Limit is enabled: check that SIL Low Alarm Limits are within approved
limits, see “Setting the SIL Low Alarm Limit” on page 29.
Manipulated Device Id, see “Device verification” on page 34.
Device Id, see “Device verification” on page 34.
Antenna Type No., Antenna Size No., Sweep Mode No., see Table 4-13 on page 36.
5. On the assumption that the device information is correct and the 5900 is properly
calibrated, the system can be set to Approved.
6. Enter the Device Id and the SIL Password. The default password=”1234”.
The password can be changed once the Safety Mode is set to “Not Approved”.
7. Click the Change to Approved button and verify that Safety Mode is changed to
“Approved”. Now the 5900 safety alarm configuration is write protected and cannot be
changed unless Safety Mode is changed to Not Approved again.
8. Verify that the Write Config Counter is incremented by one when the Safety Mode is
changed to Approved.
9. You may now click the Print Screen button to print a copy of the current Change Safety
Alarm Mode window for future reference.
10. Close the Change Safety Alarm Mode window and return to the Safety Alarm window.