Implementing a Universal Transmitter
March 2016
Reference Manual
00809-0300-4750, Rev CA
228
Implementing a Universal Transmitter
D.3.2 Model 7400 Sensor to Rosemount 8750W Transmitter
Connect coil drive and electrode cables as shown in Figure D-3.
Figure D-3. Wiring for Brooks Sensor Model 7400 and Rosemount 8750W
A. Brooks model 7400
B. Rosemount 8750W Transmitter
Refer to Figure D-1 on page 225 for actual terminal block configuration drawing.
Table D-4. Brooks Model 7400 Sensor Wiring Connections
Rosemount 8750W Brooks Sensors model 7400
1 Coils +
2 Coils –
3 3
17 Shield
18 Electrode +
19 Electrode –
Do not connect mains or line power to the magnetic flowtube sensor or to the
transmitter coil excitation circuit.