Emerson Process Management GmbH & Co. OHG 2-19
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
2
Installation
All signal cables are connected to screw-type
terminals located inside the housing. Access
to the internal components is gained by re-
leasing the two fasteners at the top of the unit
and opening the front panel downwards.
All cables must be fed through cable glands
and secured with a gland nut.
Preparation of signal cables
Gas connections
2.6.2 Installation - X-STREAM XEXF Field Housings
Installing cable glands with shielded cables
3. Feed cable through
dome nut and clamp-
ing insert.
4. Fold braided shield
over clamping insert.
5. Make sure that braid-
ed shield overlaps the
O-ring by 2 mm (
3
32“).
1. Strip the cable.
2. Expose braided
shield.
6. Push clamping insert
into body and tighten
dome nut.
7.
Assemble into hous-
ing and you´re done!
Gas connections
max number 8
max for purging (incl. / separate) 2 incl.
material PVDF; stainless steel (opt.)
sizes 6/4 mm;
1
4"
Signal in- and outputs
Properly installed, the glands act as a strain
relief and guarantee EMC (electromagnetic
compatibility):