EasyManua.ls Logo

EMS AIR-FLOW MASTER - Cleaning, Disinfecting and Sterilizing; Cleaning the Unit

EMS AIR-FLOW MASTER
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
Cleaning the unit
Clean the unit only with 70% isopropyl alcohol or other EPA recognized surface disinfectant,
following carefully the instructions provided by the disinfection solution manufacturer. The use of scouring
powder or an abrasive sponge will damage its surface.
Never remove the silicone cap from the powder chamber holder when cleaning the unit.
Do not use needles on the silicone cap of the powder chamber holder.
Please conform to the recommendations of the Reprocessing Instructions chapter of this manual.
Cleaning must be carried out immediately, no later than 30 minutes after the end of the treatment.
Follow present-day regulations enforced in the country about reprocessing.
CLEANING, DISINFECTING AND STERILIZING
FB-580_HF_Rev A-01_Not Util AIR-FLOW MASTER.indd 56 07.02.2018 14:50:22

Table of Contents

Related product manuals