98
Symptom Solution
The unit does not start or stop Check the electrical connections and power supply at the mains
Check the fuses on back panel
Check main switch function: at least one LED must be visible on adjustments and working mode areas
The unit is in "standby" mode, tap on any sensitive zone to exit standby
Send the unit to an authorized Hu-Friedy & EMS Repair Center
The foot pedal does not activate the unit Check that the foot pedal is properly connected
Replace the foot pedal
Send the unit to an authorized Hu-Friedy & EMS Repair Center
The unit is still activated when the foot
pedal is released
The foot pedal is active, check for jam at foot pedal
The switch in the pedal is blocked. Turn the unit off at the power switch
Replace the foot pedal
Send the unit to an authorized Hu-Friedy & EMS Repair Center
User interface displays an error: there
is one acoustic signal and all LEDs are
blinking
Powder chamber pressure is too low or too high
The air or liquid regulator is switched off
Foot pedal and purge commands are not recognized by system. Check connections
The handpiece does not remain on the
magnetic support
Install the magnetic support on a metallic part of the unit
The support is demagnetized
Replace magnetic support
The handpiece does not hold on the
handpiece cord
Check for presence of the O-ring seal and replace if necessary
Replace the handpiece
Replace the handpiece cord
Send the handpiece and the handpiece cord to an authorized Hu-Friedy & EMS Repair Center
No water jet coming from the handpiece Check the water supply and water lter
Check the water adjustment of your unit
Check the connections between the unit, the cord and the handpiece
Check for presence of water at the cord outlet port
Send the handpiece to an authorized Hu-Friedy & EMS Repair Center
No powder/air jet coming from the
handpiece
Check the quantity of powder in your unit
Check the compressed air supply and air lter
Check for presence of powder/air at cord outlet port
Eliminate residual powder present in the channels using the cleaning needles provided
AIR-FLOW PERIO+ PLUS powder chamber only: remove the diffuser and empty the powder. Blow the diffuser with
compressed air before installing it again and relling.
Send the system to an authorized Hu-Friedy & EMS Repair Center
TROUBLESHOOTING
Cont. next page