For Endochuck file holder:
Reinstall the small O-ring (gasket) first, and then gently screw the retention bolt without
tightening it.
For any Instrument having its CombiTorque:
Reinstall the Instrument on the CombiTorque.
Any EMS part can be correctly packaged using the following validated procedure:
Single or Double Pouches suitable for Pre-Vacuum Moist Heat Sterilization compliant with
ISO 11607-1, or EN 868 resistant to 138°C and having adequate steam permeability (e.g.
Wipak STERIKING flat rolls Type R43 and R44).
Sterilization
Sterilization must be performed immediately after cleaning-disinfection.
DO NOT exceed the maximum number of sterilization cycles allowed.
DO NOT exceed a sterilization temperature of 138°C and a holding time of 20 minutes.
DO NOT use hot-air sterilization and radio-sterilization procedures: they destroy the products.
Moist heat sterilization of parts shall be performed according to ISO 17665 and under consideration
of the respective country requirements.
The following validated Pre-vacuum Moist Heat (steam) process can be used with any EMS part
packaged in appropriate single or double pouches:
Parameters for the Pre-vacuum Moist Heat cycle:
3 pre-vacuum phases
Sterilization temperature of 132°C
Pressure of 3 bar (absolute pressure)
Humidity of 100%
Holding time of 3 minutes minimum (full cycle)
Drying time of 20 minutes minimum

Other manuals for EMS AIRFLOW Prophylaxis Master