35
Figure 87
1. Press the highlighted faulty component and follow
the interactive menu to identify the exact origin of
the error.
2. Follow the recommended action that is displayed.
3. If the solutions proposed fail to solve the problem,
please contact your EMS authorized service center.
Do not, in any case, return a product before trouble-
shooting of the error has been performed.
4. The following table provides more detailed information
about failures: error number and associated error
messages.
Console
E001 - The cooling pump is not detected and handpiece cooling might not be available. Please restart device.
Please contact your EMS authorized service center if the error persists.
E002 - The cooling valve is not detected and handpiece cooling might not be available. Please restart device.
Please contact your EMS authorized service center if the error persists.
E008 - Conguration les of the console are corrupted and informations might be incorrect. Please contact your
EMS authorized service center.
E009 - Console internal communication error. Please restart device. Please contact your EMS authorized service
center if the error persists.
E010 - Pedal not detected. Please verify that the connector of the pedal is connected to the console. Please
contact your EMS authorized service center if the error persists.
E016 - No suction system has been detected. Please restart device or contact your EMS authorized service center
if the error persists.
E017 - Two suction systems seem to be connected. Please restart device or contact your EMS authorized service
center if the problem persists.
E018 - The console temperature is high. Treatment is still possible but verify the console is placed in a correctly
ventilated place
E019 - The console temperature is too high. System needs to cool down. Please keep it powered while temperature
returns to safe level.
E020 - Console internal communication error. System trying to recover. Please restart device or contact your EMS
authorized service center if the error persists.
E024 - Console internal communication error. Please restart device or contact your EMS authorized service center
if the error persists.