Commissioning Deltabar FMD72
36 Endress+Hauser
Description
5 Select the "Dry" option via the
"Calibration mode (027)" parameter.
Menu path: Setup ® Extended Setup
® Level ® Calibration Mode
1000
0
V
[l]
50 450
p
[mbar]
20
4
I
[mA]
0 1000
V
[l]
A0017662
1
See table, step 6.
2 See table, step 7.
3 See table, step 8.
4 See table, step 9.
5 See table, step 11.
6 See table, step 12.
6 Enter the volume value for the lower
calibration point via the "Empty Calib.
(028)" parameter, here 0 liters for
example.
Menu path: Setup ® Extended Setup
® Level ® Empty Calibration
7 Enter the pressure value for the lower
calibration point via the "Empty Pressure
(029)" parameter, here 50 mbar (0.75
psi), for example.
Menu path: Setup ® Extended Setup
® Level ® Empty Pressure
8 Enter the volume value for the upper
calibration point via the "Full Calib.
(031)" parameter, here 1000 liters (264
gal) for example.
Menu path: Setup ® Extended Setup
® Level ® Full Calibration
9 Enter the pressure value for the upper
calibration point via the "Full Pressure
(032)" parameter, here 450 mbar (6.75
psi), for example.
Menu path: Setup ® Extended Setup
® Level ® Full Pressure
10 "Adjust density (034)" contains the
factory setting 1.0, but this value can be
changed if required.
Menu path: Setup ® Extended Setup
® Level ® Adjust Density
11 Set the volume value for the lower
current value (4 mA) using the "Set LRV
(166)" parameter (0 l).
Menu path: Setup ® Extended Setup
® Current Output ® Set LRV
12 Set the volume value for the upper
current value (20 mA) using the "Set
URV (167)" parameter (1000 l (264 gal)).
Menu path: Setup ® Extended Setup
® Current Output ® Set URV