Micropilot S FMR533 4 to 20 mA HART Commissioning
Endress+Hauser 43
6.4 Basic Setup with the VU331
Function "measured value" (000)
This function displays the current measured value in the selected unit
(see "customer unit" (042)) function). The number of digits after decimal point can be selected in
the "no.of decimals" (095) function. The length of the bargraph corresponds to the percental
value of the present measured value with regard to the span.
6.4.1 Function group "basic setup" (00)
Function "tank shape" (002)
This function is used to select the tank shape.
Selection:
dome ceiling
horizontal cyl
bypass (Not weights and measures approved, accuracy is not guaranteed.
Recommendation: FMR532).
stilling well (Not weights and measures approved, accuracy is not guaranteed.
Recommendation: FMR532).
flat ceiling (Typical ceiling of storage tanks: a slight slope of only a few degrees can be neglected.)
•sphere
L00-FMR2xxxx-14-00-06-en-007
ENDRESS + HAUSER
E
+
ENDRESS + HAUSER
E
+
ENDRESS + HAUSER
E
+
dome ceiling
flat ceiling sphere
horizontal cyl
bypass stilling well

Table of Contents