Commissioning Micropilot S FMR533 4 to 20 mA HART
64 Endress+Hauser
If "flat ceiling" is selected in the "tank shape" function, the following display appears on the
screen:
• empty calibr.
• full calibr.
MicropilotS-en-006
Basic Setup step 4/5:
• This step starts the tank mapping
• The measured distance and the current measured value are always displayed in the header
• A description is given, → ä 48
MicropilotS-en-007
Basic Setup step 5/5:
After the first installation of the device, initialise the index correction table. To do so set the history
reset to "yes".