Levelflex FMP51, FMP52, FMP54
Endress+Hauser 113
Type of probe Minimum distance from
end of probe to 0% level
Maximum value for
"empty calibration"
Rod (not divisible) C ≥ 100 mm (4 in) E ≤ 3.9 m (12.8 ft)
Coax
Rod (divisible)
C ≥ 100 mm (4 in) E ≤ 5.9 m (19.4 ft)
Rope C ≥ 1 000 mm (40 in) E ≤ 23 m (75 ft)
For rod and rope probes the linearity check is performed with the complete device.
For coax probes the electronics is mounted to a reference rod probe and the linearity check is
performed to this configuration.
The linearity is checked under reference conditions.
The selected values of Empty calibration and Full calibration are only used to record the
linearity protocol and are reset to their probe specific default values thereafter. If values
different from the default are required, they must be ordered as a customized parametrization
→  114.