Applications with restricted mounting space
Mounting with remote sensor
The device version with a remote sensor is suited for applications with restricted mounting space. In
this case the electronics housing is mounted at a separate position from which it is easier accessible.
A
B
C
C
r = 100 (4)
min
r = 100 (4)
min
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
A0014794
A Angled plug at the probe
B Angled plug at the electronics housing
C Length of the remote cable as ordered
Product structure, feature 600 "Probe Design":
Option MB "Sensor remote, 3m/9ft cable"
The remote cable is supplied with these device versions
Minimum bending radius: 100 mm (4 inch)
A mounting bracket for the electronics housing is supplied with these device versions. Mounting
options:
Wall mounting
Pipe mounting; diameter: 42 to 60 mm (1-1/4 to 2 inch)
The connection cable has got one straight and one angled plug (90°). Depending on the local
conditions the angled plug can be connected at the probe or at the electronics housing.
Probe, electronics and connection cable are adjusted to match each other. They are marked by a
common serial number. Only components with the same serial number shall be connected to
each other.