EasyManua.ls Logo

Endress+Hauser Levelflex M FMP40 - Mounting probes in an empty silo

Endress+Hauser Levelflex M FMP40
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting Levelflex M FMP40 FOUNDATION Fieldbus
16 Endress+Hauser
Mounting Location
Recommended distance B wall-mounted
rope probe: ~1/6 to 1/4 of the container
diameter.
Not central (2) in metallic tanks.
Not in the filling curtain (3).
Please order the probe length such that it
ends approx 30 mm above the floor of the
tank.
Temperature conditions must be met.
It is recommended that a protective cover
(1) be used, in order to protect the
transmitter against direct sunlight or rain.
Mounting and demounting are carried out
simply with a clamp (see Operating
Instruction on the CD-ROM).
L00-FMP4xxxx-17-00-00-xx-001
B
123

Other manuals for Endress+Hauser Levelflex M FMP40