Levelflex M FMP40 and FMP45 Installation
15
3.1.2 FMP45
L00-FMP45xxx-17-00-00-en-003
3.2 Incoming acceptance, transport, storage
3.2.1 Incoming acceptance
Check the packing and contents for any signs of damage.
Check the shipment, make sure nothing is missing and that the scope of supply matches your order.
3.2.2 Transport
"
Caution!
Follow the safety instructions and transport conditions for instruments of more than
18 kg.
Do not lift the measuring instrument by its probe rod in order to transport it.
3.2.3 Storage
Pack the measuring instrument so that is protected against impacts for storage and transport. The
original packing material provides the optimum protection for this.
The permissible storage temperature is -40 °C…+80 °C.
3
1
2
60
1"
½
F12 housing
Turn housing
The housing can be turned 350°
in order to simplify access to the
display and the terminal compartment
AF 60
max. torque
see table
Caution!
Use only the
threaded boss ???
F12 or T12 housing
torqueProcess temperature
max. 280 °C 450 Nm
max. 400 °C 400 Nm