7 Description of Instrument Functions Levelflex M FMP40 and FMP45
50 Endress+Hauser
sim. level
Enter the level value in "simulation value".
The functions
•measuredvalue
• measured level
• output current" - only with HART instruments!
follow the entered values.
sim. volume
Enter the volume value in "simulation value".
The functions
•measuredvalue
• output current" - only with HART instruments!
follow the entered values.
sim. current (HART only)
Enter the current value in "simulation value".
The function
• output current" - only with HART instruments!
follows the entered values.
7.5.7 Function "simulation value"
7.5.8 Function "output current", HART only
Displays the output current in mA.
"
Caution!
This function is available for HART devices only!
7.5.9 Function "4mA value", HART only
In this function, enter the level (or volume, weight) at which the output current should be 4 mA.
This entry is only required if you selected the
"
current turn down
"
option in the
"
curr. output
mode
"
function.
7.5.10 Function "20mA value", HART only
In this function, enter the level (or volume, weight) at which the output current should be 20 mA.
This entry is only required if you have selected the
"
current turn down
"
option in the
"
curr. output mode
"
function.
7.6 Function group "display"
7.6.1 Function "language"
Selects the display language.
Selection:
•English
•Deutsch
• Français
•Español
• Italiano
• Nederlands
• Katakana (japanese)