EasyManua.ls Logo

Endress+Hauser Levelflex M FMP45 - Page 27

Endress+Hauser Levelflex M FMP45
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex M FMP40 and FMP45 Wiring
27
4.2.1 HART connection with E+H RMA 422 / RN 221 N
L00-FMP4xxxx-04-00-00-yy-001
4.2.2 HART connection with other supplies
L00-FMP4xxxx-04-00-00-en-002
"
Caution!
If the HART communication resistor is not built into the supply unit and the HART interface is to
be used, it is necessary to insert a communication resistor of 250 into the 2-wire line.
4...20 mA
1
2
3
4
ENDRESS + HAUSER
RMA 422
Commubox
FXA191
RMA422
RN221N
- FieldCare
1
2
3
4
- FieldCare
Commubox
FXA191
³W250
PLC
DC power supply
unit
or
4...20 mA