7 Description of Instrument Functions Levelflex M FMP40 and FMP45
42 Endress+Hauser
If the instrument is in alarm state, the output changes as follows:
HART: MAX-Alarm 22 mA
hold
If the instrument is in alarm state, the last measured value is held.
user specific
If the instrument is in alarm state, the output is set to the value configured in "output on alarm"
(x mA).
"
Caution!
This selection is available for HART devices only!
7.2.2 Function "output on alarm", HART only
On alarm, the output current is in mA. This function is active when you selected "user specific"
in the "output on alarm" function.
"
Caution!
This function is available for HART devices only!