Micropilot FMR10
18 Endress+Hauser
Environment
Ambient temperature range
Measuring device: –40 to +60 °C (–40 to +140 °F)
For outdoor operation in strong sunlight:
• Mount the device in the shade.
• Avoid direct sunlight, particularly in warm climatic regions.
• Use a weather protection cover.
Storage temperature
–40 to +80 °C (–40 to +176 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Operating altitude as per
IEC 61010-1 Ed.3
Generally up to 2 000 m (6 600 ft) above sea level.
Degree of protection
Tested acc. to:
• IP66, NEMA 4X
• IP68, NEMA 6P (24 h at 1.83 m (6.00 ft)1.83 m under water)
Vibration resistance
DIN EN 60068-2-64/IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s
2
)
2
/Hz
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61000 series and NAMUR Recommendation EMC (NE 21). Details are provided in the
Declaration of Conformity (www.endress.com/downloads).