What to do if Endress+Hauser Radar measured value jumps sporadically to a higher level?
- JJeremy MillerAug 27, 2025
If the measured value jumps sporadically to a higher level when the surface is not calm (e.g., during filling, emptying, or when an agitator is running) with your Endress+Hauser Radar, it's likely because the signal is weakened by the rough surface, making interference echoes stronger. Here are some steps to take: * Carry out tank mapping. * Select 'Process vessel with agitator' for the Tank type. * Increase the integration time. * Optimize the antenna orientation. * Select a better mounting position and/or larger antenna if necessary.