Promag P tightening torques for JIS
Promag P tightening torques for AS 2129
Promag P tightening torques for AS 4087
Nominal diameter JIS Max. tightening torque [Nm]
[mm] Pressure rating Threaded fasteners PTFE PFA
25 10K 4 × M 16 32 27
25 20K 4 × M 16 32 27
32 10K 4 × M 16 38
32 20K 4 × M 16 38
40 10K 4 × M 16 41 37
40 20K 4 × M 16 41 37
50 10K 4 × M 16 54 46
50 20K 8 × M 16 27 23
65 10K 4 × M 16 74 63
65 20K 8 × M 16 37 31
80 10K 8 × M 16 38 32
80 20K 8 × M 20 57 46
100 10K 8 × M 16 47 38
100 20K 8 × M 20 75 58
125 10K 8 × M 20 80 66
125 20K 8 × M 22 121 103
150 10K 8 × M 20 99 81
150 20K 12 × M 22 108 72
200 10K 12 × M 20 82 54
200 20K 12 × M 22 121 88
250 10K 12 × M 22 133
250 20K 12 × M 24 212
300 10K 16 × M 22 99
300 20K 16 × M 24 183
Nominal diameter
[mm]
AS 2129
Pressure rating
Threaded fasteners Max. tightening torque [Nm]
PTFE
25 Table E 4 × M 12 21
50 Table E 4 × M 16 42
Nominal diameter
[mm]
AS 4087
Pressure rating
Threaded
fasteners
Max. tightening torque [Nm]
PTFE
50 PN 16 4 × M 16 42