Proline Prosonic Flow 93 Installation
Endress+Hauser
7
A0001105
Fig. 1: Horizontal orientation
A Recommended orientation with upward direction of flow
B Recommended installation range with horizontal orientation
C Recommended installation range max. 120°
Inlet and outlet run
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following inlet and outlet runs is recommended in order to ensure
measuring accuracy.
A0013459
Fig. 2: Inlet and outlet run (top view)
1 Valve (2/3 open)
2 Pump
3 Two pipe bends in different directions
A
B
CC
³ 15xDN
³ 20xDN
³ 20xDN
³ 15xDN
³ 3xDN
1
2
3