Installation Proline Prosonic Flow 93
12 Endress+Hauser
3.6 Determining the installation distances
3.6.1 Running the "Sensor Installation" Quick Setup menu
Calling the Quick Setup
All the installation distances required for sensor mounting are determined with the Quick Setup.
!
Note!
If you are not familiar with the operation of the device ä 36.
1. F Enter the function matrix (starting with measured value display)
2. P Select the group QUICK SETUP
F Confirm selection
3. P Select the SENSOR INSTALLATION function
F Confirm selection
4. Intermediate step if configuration is blocked:
P Enter the code 93 (confirm with F ) and thus enable configuration
5. P Go to Sensor Installation Quick Setup
6. P Select YES
F Confirm selection
7. F Start Sensor Installation Quick Setup
For subsequent steps, see the following section "Running the Quick Setup for Clamp-on type of
mounting".

Other manuals for Endress+Hauser Proline Prosonic Flow 93

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 93 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 93 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 93
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals