Installation Proline Prosonic Flow 93
14 Endress+Hauser
3.7 Mechanical preparation
The way in which the sensors are secured differs on account of the pipe nominal diameter and
the sensor type. Depending on the type of sensor, users also have the option of securing the
sensors with strapping bands or screws such that they can be later removed, or permanently
fixing the sensors in place with welded bolts or welded holders.
Overview of possible ways to secure the various sensors:
3.7.1 Mounting the holder with U-shaped screws
Sensor: Prosonic Flow 93P and Prosonic Flow 93W (DN 15 to 65 / ½ to 2½")
For mounting on a pipe with a nominal diameter of DN 32 (1¼").
1. Disconnect the sensor from the holder.
2. Position the holder on the pipe.
3. Put the U-shaped screws through the holder and slightly lubricate the thread.
4. Screw nuts onto the U-shaped screws.
5. Set the holder to the exact position and tighten the nuts evenly.
#
Warning!
Risk of damaging plastic or glass pipes if the nuts of the U-shaped screws are tightened too
much! The use of a metal half-shell is recommended (on the opposite side of the sensor)
when working with plastic or glass pipes.
A0001109
Fig. 3: Mounting the Prosonic Flow P holder (DN 15 to 65 / ½ to 2½") with U-shaped screws
Prosonic Flow Pipe nominal diameter Secured by
93P
93W
DN 15 to 65
(½ to 2½")
DN 32 (1¼") Holder with U-shaped screws ä 14
DN 32 (1¼") Holder with strapping bands ä 15
93P
93W
DN 50 to 4000
(2 to 160")
DN 200 (8") Strapping bands (medium nominal diameters) ä 16
Welded bolts ä 18
DN 200 (8") Strapping bands (large nominal diameters) ä 17
Welded bolts ä 18