Installation Proline Prosonic Flow 93
10 Endress+Hauser
3.4.3 Panel mounting
A0007544
3.5 Connecting the power supply
#
Warning!
Risk of electric shock. Components carry dangerous voltages.
Never mount or wire the measuring device while it is connected to the power supply.
Before connecting the power supply, check the safety equipment.
Route the power supply cable so that it is seated securely.
Seal the cable entries and covers tight.
When wiring Ex-certified measuring devices, all the safety instructions, wiring diagrams,
technical information, etc. of the related Ex documentation must be observed
Ex documentation on the CD-ROM.
"
Caution!
Risk of damaging the electronic components!
Connect the power supply in accordance with the connection data on the nameplate.
3.5.1 Connecting the transmitter
Wire the unit using the terminal assignment diagram inside the cover.
245 (9.65)
~110 (~4.33)
210 (8.27)
+0.5 (+0.019)
–0.5 (–0.019)
+0.5 (+0.019)
–0.5 (–0.019)
mm (inch)
A0010773
Transmitter connection:
1
2
Connection compartment cover
(connection diagram on the inside)
Power supply cable
2
1

Other manuals for Endress+Hauser Proline Prosonic Flow 93

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 93 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 93 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 93
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals