Example for max. measured error (volume flow)
0
[%]
2.0
1.0
0
5
v
[m/s]
[ft/s]
3.0
4.0
30
0
5 10 15 100
5.0
6.0
7.0
8.0
9.0
10.0
2
1
20
25
1 2 3 4
6
7 8
A0015541
35 Example for max. measured error (volume flow) in % o.r.
1 Standard (order code for "Calibration flow", option 1 "Operable flow range 30 : 1")
2 Optional (order code for "Calibration flow", option 2 "Operable flow range 100 : 1")
Accuracy of outputs
o.r. = of reading
The outputs have the following base accuracy specifications.
Current output
Accuracy ±10 µA
Pulse/frequency output
Accuracy Max. ±100 ppm o.r.
Repeatability o.r. = of reading; o.f.s. = of full scale value; abs. = absolute; T = medium temperature
Volume flow
±0.5 % o.r.
Methane
±0.5 % o.f.s. = ±0.5 % abs.
Temperature
±0.3 °C ± 0.0025 × T °C (±0.45 °F ± 0.0025 × (T – 32) °F)
Response time • The response time depends on the configuration (damping).
• Response time in the event of erratic changes in the flow: after 1 000 ms 95% of the full
scale value.
Influence of ambient
temperature
o.r. = of reading