Technical data Proline Prosonic Flow B 200 HART
142 Endress+Hauser
Cable entries/cable glands
1
2
3
A0020640
 36 Possible cable entries/cable glands
1 Cable entry in transmitter housing or connection housing with internal thread M20 x 1.5
2 Cable gland M20 x 1.5
3 Adapter for cable entry with internal thread G ½" or NPT ½"
Order code for "Housing", option C "GT20 two-chamber, aluminum coated"
Transmitter
Cable entry/cable gland Type of protection Material
Cable gland M20 × 1.5
Non-Ex
Ex ia
Plastic
Adapter for cable entry with
internal thread G ½"
For non-Ex and Ex
(except for CSA Ex d/XP)
Nickel-plated brass
Adapter for cable entry with
internal thread NPT ½"
For non-Ex and Ex
Transmitter neck
Cable gland Measuring path Material
Cable gland M20 × 1.5 Two-path Nickel-plated brass
Cable gland M12 × 1.5 Single-path
Sensor
Cable gland Material
Cable gland M12 × 1.5 Nickel-plated brass
Order code for "Housing", option S, "GT18 two-chamber, stainless steel"
Transmitter
Cable entry/cable gland Type of protection Material
Cable gland M20 × 1.5 Non-Ex
Ex ia
Stainless steel ,1.4404
Adapter for cable entry with
internal thread G ½"
For non-Ex and Ex
(except for CSA Ex d/XP)
Stainless steel, 1.4404 (316L)
Adapter for cable entry with
internal thread NPT ½"
For non-Ex and Ex