Proline Prowirl R 200 Installation
Endress+Hauser 17
6.1.3 Special mounting instructions
Outlet runs for pressure and temperature measuring points
If pressure and temperature measuring points are installed after the device, please
ensure there is a large enough distance between the device and the measuring point so
there are no negative effects on vortex formation in the sensor.
PT
TT
3...5 x DN
4...8 x DN
A0019205
PT Pressure measuring point
TT Temperature measuring point
Open aperture flow conditioner
If the required inlet runs cannot be observed, it is possible to install a specially designed
open aperture flow conditioner which can be ordered from Endress+Hauser. The flow
conditioner is fitted between two pipe flanges and centered by the mounting bolts.
Generally this reduces the inlet run needed to
10 × DN with full accuracy.
8 x DN
2 x DN
5 x DN
1
A0019208
1 Flow conditioner