3.2 Transport instructions for Promag from > DN 350 / 14"
The pipe lining on the flanges is protected by disks to prevent damage when
transporting to the installation point. These are to be removed before installation.
The instruments are to be transported in the containers they are delivered in.
Transporting to the measuring point
The grips on the flange must be used
for lifting out and mounting the sensor
in the piping.
The sensor should not be lifted by the
connection housing.
The sensor should not be lifted by the
sheet casing using a forklift truck.
This can buckle the casing and
damage the internal magnetic coils.
Base of the sensor
The sensor should stand on a base
which is sufficiently strong to withstand
its weight.
Note!
The sensor must not be supported by
the sheet casing! This can buckle the
casing and damage the internal
magnetic coils.
ba008y60
ba008y61
ba008y62
Fig. 6
ba008y63
ba008y64
Fig. 7
3. Mounting and Installation Promag 30
14 Endress+Hauser