Conversion transmitter Promag 50/53 W, P, H to Promag 55 remote version
Endress + Hauser 3
A Information about the modified device
Relevant documentation
Not only the already existing documentation, but also the documentation delivered with the transmitter will be relevant additionally,
be
cause of the conversion.
For the sensor the documentation for Promag 50, 53 is relevant.
For the transmitter the documentation for Promag 55 is relevant.
Changes and restrictions conditioned by conversion
Accuracy
After conversion of the device an additional measured error of ± 0,2% to the measured value must be calculated.
B Preparatory operations: Save device data
Conversion Promag 50
Select important parameter out of the application (e.g. with the aid of the Quick Setup "Commisioning" see Operating Instructions) and
fill them into the "list of parameter", chapter H. After Conversion enter the noted parameter into the new transmitter.
Conversion Promag 53
Save the parameters of the transmitter to the T-DAT using the "T-DAT SAVE/ LOAD" function".
Replace the T-DAT into the transmitter after conversion.
C Overview data storage (Histo-ROM) and F-CHIP
Data storage Colour Promag 50 to Promag 55 Promag 53 to Promag 55
S-DAT blue Re-assembly Re-assembly
T-DAT black New Re-assembly
F-CHIP (Optional) red New New

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Promag Series and is the answer not in the manual?

Endress+Hauser Promag Series Specifications

General IconGeneral
Measurement PrincipleElectromagnetic
Output4 to 20 mA
Communication ProtocolsHART, PROFIBUS DP, FOUNDATION Fieldbus, Modbus
Enclosure RatingIP67, NEMA 4X
DisplayLCD with backlight
Housing MaterialAluminum, stainless steel
Ambient Temperature-40°C
Repeatability±0.1% of reading
CommunicationHART, PROFIBUS DP, FOUNDATION Fieldbus, Modbus

Related product manuals