Prosonic S FDU90/91/91F/92/93/95/96
12 Endress + Hauser
Installation conditions for flow
measurements
Install the sensor at the inflow side (A), above the maximum water level H
max
plus the blocking distance BD.
Position the sensor in the middle of the channel or weir.
Align the sensor vertically to the water surface.
Comply to the installation distance of the channel or weir.
2)
Use a protective cover, in order to protect the sensor from direct sun or rain. A protective cover is available
for the sensors FDU90 and FDU91 (see chapter "Accessories").
Example: Khafagi-Venturi flume
L00-FDU9xxxx-17-00-00-xx-004
(A): inflow side; (B): outflow side
Example: V-notch weir
L00-FDU9xxxx-17-00-00-xx-005
2) The installation distances of important flumes and weirs are specified in the Operating Instructions BA 289F (FMU90 with HART) and BA 293F (FMU90 with
PROFIBUS).
1 x b
0
b
0
BD
FDU 91
H
max
(A) (B)
max
min. 3 H
BD
FDU 91
H
max