Technical Data Prosonic T
48 Endress+Hauser
10 Technical Data
10.1 Technical data at a glance
10.1.1 Input
Measured variable The distance D between the sensor membrane and the product surface is measured.
Using the linearisation function, the device uses D to calculate:
• level L in any units
• volume V in any units
• flow Q across measuring weirs or open channels in any units
Maximum range/blocking
distance
10.1.2 Output
Output signal 4 to 20 mA
Signal on alarm • Error symbol, error code and plain text description on the on-site display
• Current output (configurable)
10.1.3 Power supply
Terminals Cable cross-section: 0.25 to 2.5 mm (20 to 14 AWG)
Cable entry • Cable gland M20x1.5 (recommended cable diameter 6 to 10 mm (0.24 to 0.39 in))
• Cable entry G½ or ½ NPT
Supply voltage 14 to 35 V
There may be additional restrictions for devices with an explosion protection certificate. Refer to the
notes in the appropriate safety instructions (XA).
Power consumption 51 mW to 800 mW
Sensor BD Max. range fluids
1)
1) The actual range is dependent on the measuring conditions. Refer to Technical Information TI00440F/00/EN for
an estimation.
Max. range bulk materials
1½" 0.25 m (0.8 ft) 5 m (16 ft) 2 m (6.6 ft)
2" 0.35 m (1.1 ft) 8 m (26 ft) 3.5 m (11 ft)