Soliwave FDR56/FQR56
12 Endress+Hauser
Mounting
4.2 Mounting the device
4.2.1 Mounting with connection thread
0000000166
7 Mounting with connection thread
A 1½NPT
B G1/G1½
Screw in connection thread. For G 1½ thread: Tighten counternut.
Align the electronics housing (cable gland or plug connector of both devices must point in
the same direction).
Fix the housing in place.