Soliwave FDR56/FQR56
18 Endress+Hauser
Electrical connection
The contact material of the relay is also suitable for switching small signal circuits, if no
inductive loads or higher currents have been switched previously.
If the switching frequency is high, the solid-state relay should be selected.
When using the Harting connector type HAN8D the maximum switching voltage is 120V
DC or 50V AC.
Current output
Current 4-20mA, active
Load max. 600Ω
Electrical connection Current output
Connection terminals Terminals 3 (+) - 4 (-)
M12 connector Binder series 713/763 Connector 2, contact 3 (+) - 4 (-)
Harting connector type HAN8D Connector 2, contact 3 (+) - 4 (-)
Solid-state relay
Switching capacity 30V AC / 0.4A or 40V DC / 0.4A
Switching delay parameterizable (o, 500ms to 10s)
Switching hysteresis adjustable
Switching frequency max. 4Hz
Electrical connection Solid-state relay
Connection terminals Terminals 3 - 4
M12 connector Binder series 713/763 Connector 2, contact 3 - 4
Harting connector type HAN8D Connector 2, contact 3 - 4
The polarity of the solid-state relay can be set as required.
5.3 Post-connection check
Is the device or cable undamaged (visual inspection)?
Do the cables used comply with the requirements?
Do the mounted cables have adequate strain relief?
Are the connectors rmly tightened?
Does the supply voltage match the specications on the nameplate?
No reverse polarity, is terminal assignment correct?
If supply voltage is present, is the green LED lit?