Turbimax CUS51D Product description
Endress+Hauser 11
S
1
E
1-135
E
2-135
S
2
E
1-90
E
2-90
A0030847
 5 4-beam pulsed light method
S
1
S
2
Light source
E
90
Light receiver for 90° scattered light
E
135
Light receiver for 135° scattered light
90° scattered light method
Measurement is performed with a wavelength of 860 nm, as described in ISO 7027/EN
27027.
The emitted light beam is scattered by the solid particles in the medium. The scattered
radiation generated in this way is measured by scattered light receivers, which are
arranged at an angle of 90° to the light sources. The turbidity of the medium is determined
by the amount of scattered light.
S
E
a
P
A0030852
 6 90° scattered light method
S Light source
E Receiver
P Particle
135° backscattered light method
The emitted light beam is scattered by the solid particles in the medium. The
backscattering generated is measured by scattered light receivers, which are arranged next
to the light sources. The turbidity of the medium is determined based on the quantity of
back-scattered light. It is possible to measure very high turbidity values with this type of
scattered light measurement.

Other manuals for Endress+Hauser Turbimax CUS51D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Turbimax CUS51D and is the answer not in the manual?

Endress+Hauser Turbimax CUS51D Specifications

General IconGeneral
BrandEndress+Hauser
ModelTurbimax CUS51D
CategoryAccessories
LanguageEnglish

Related product manuals