Turbimax CUS51D Maintenance
Endress+Hauser 39
10 Maintenance
L
CAUTION
Acid or medium
Risk of injury, damage to clothing and the system!
‣
Switch off cleaning before the sensor is removed from the medium.
‣
Wear protective goggles and safety gloves.
‣
Clean away splashes on clothes and other objects.
‣
You must perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the following:
• The system
• The installation conditions
• The medium in which measurement takes place
10.1 Maintenance tasks
NOTICE
Disassembly at sensor head
Sensor can leak!
‣
Turn by the shaft only.
‣
Never turn by the sensor head!
10.1.1 Cleaning the sensor
Sensor fouling can affect the measurement results and even cause a malfunction.
‣
To ensure reliable measurements, clean the sensor at regular intervals. The frequency
and intensity of the cleaning depend on the medium.
Clean the sensor:
• As specified in the maintenance schedule
• Before every calibration
• Before returning it for repairs
Type of fouling Cleaning measure
Lime deposits
‣
Immerse the sensor in 1 to 5% hydrochloric acid (for several minutes).
Dirt particles on the optics
‣
Clean the optics with a cleaning cloth.
After cleaning:
‣
Rinse the sensor thoroughly with water.