Turbimax CUS51D Maintenance
Endress+Hauser 39
10 Maintenance
L
CAUTION
Acid or medium
Risk of injury, damage to clothing and the system!
Switch off cleaning before the sensor is removed from the medium.
Wear protective goggles and safety gloves.
Clean away splashes on clothes and other objects.
You must perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the following:
The system
The installation conditions
The medium in which measurement takes place
10.1 Maintenance tasks
NOTICE
Disassembly at sensor head
Sensor can leak!
Turn by the shaft only.
Never turn by the sensor head!
10.1.1 Cleaning the sensor
Sensor fouling can affect the measurement results and even cause a malfunction.
To ensure reliable measurements, clean the sensor at regular intervals. The frequency
and intensity of the cleaning depend on the medium.
Clean the sensor:
As specified in the maintenance schedule
Before every calibration
Before returning it for repairs
Type of fouling Cleaning measure
Lime deposits
Immerse the sensor in 1 to 5% hydrochloric acid (for several minutes).
Dirt particles on the optics
Clean the optics with a cleaning cloth.
After cleaning:
Rinse the sensor thoroughly with water.