Turbimax CUS51D Electrical connection
Endress+Hauser 25
Cable sample (does not necessarily correspond to the original cable supplied)
1
2
3
 24 Terminated cable
1 Outer shield (exposed)
2 Cable cores with ferrules
3 Cable sheath (insulation)
4
A0045763
 25 Connect the cable to the
grounding clamp
4 Grounding clamp
A0045764
 26 Press the cable into the
grounding clamp
The cable shield is grounded using
the grounding clamp
1)
1) Please note the instructions in the "Ensuring the degree of protection" section
1. Loosen a suitable cable gland on the bottom of the housing.
2. Remove the dummy plug.
3. Attach the gland to the cable end, making sure the gland is facing the right direction.
4. Pull the cable through the gland and into the housing.
5. Route the cable in the housing in such a way that the exposed cable shield fits into
one of the cable clamps and the cable cores can be easily routed as far as the
connection plug on the electronics module.
6. Connect the cable to the cable clamp.
7. Clamp the cable.
8. Connect cable cores as per the wiring diagram.
9. Tighten the cable gland from outside.
6.2 Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these instructions
and which are necessary for the required, designated use, may be carried out on the device
delivered.
Exercise care when carrying out the work.
Individual types of protection permitted for this product (impermeability (IP), electrical
safety, EMC interference immunity) can no longer be guaranteed if, for example :
Covers are left off
Different power units to the ones supplied are used
Cable glands are not sufficiently tightened (must be tightened with 2 Nm (1.5 lbf ft) for
the permitted level of IP protection)
Unsuitable cable diameters are used for the cable glands
Modules are not fully secured
The display is not fully secured (risk of moisture entering due to inadequate sealing)
Loose or insufficiently tightened cables/cable ends
Conductive cable strands are left in the device