Electrical connection Turbimax CUS51D
24 Endress+Hauser
6 Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣
Prior to commencing connection work, ensure that no voltage is present on any cable.
6.1 Connecting the sensor
The following connection options are available:
• Via M12 plug (version: fixed cable, M12 plug)
• Via sensor cable to the plug-in terminals of a sensor input on the transmitter (version:
fixed cable, end sleeves)
85 86
85
2DS
1
2
86
97 88 8798
97 88 8798
Sensor 1
Sensor 2
PK
GY
GN
YE
Sensor
85 86
85
1
2
86
97 88 8798
Sensor 1
PK
GY
GN
YE
BN
WH
Sensor
A0033092
23 Sensor connection to sensor input (left) or via M12 plug (right)
The maximum cable length is 100 m (328.1 ft).
6.1.1 Connecting the cable shield
Device cable must be shielded cables.
Only use terminated original cables where possible.
Clamping range of cable clamps: 4 to 11 mm (0.16 to 0.43 in)