Turbimax CUS51D Technical data
Endress+Hauser 45
Electromagnetic
compatibility (EMC)
Interference emission and interference immunity according to:
• EN 61326-1:2013
• EN 61326-2-3:2013
• NAMUR NE21: 2012
13.5 Process
Process temperature range –5 to 50 °C (23 to 122 °F)
Up to 80 °C (176 °F) for a short period of time (1 h)
Process pressure range 0.5 to 10 bar (7.3 to 145 psi) (abs.)
Compressed air cleaning
Pressure: 1.5 to 2 bar (21.8 to 29 psi)
Minimum flow No minimum flow required.
For solids which have a tendency to form deposits, ensure that sufficient mixing is
performed.
13.6 Mechanical construction
Dimensions → Section "Installation"
Weight Approx. 0.7 kg (1.5 lb)without cable
Materials Sensor Stainless steel 1.4404 (AISI 316 L)
Stainless steel 1.4571 (AISI 316 Ti)
Optical windows Sapphire
O-rings EPDM
Process connections G1 and NPT ¾'
Compressed air cleaning
6 mm (0.24 in) or 8 mm (0.31 in) or 6.35 mm (0.25 in) (¼")