Electrical connection Turbimax CUS51D
26 Endress+Hauser
6.3 Post-connection check
Device health and specifications Action
Is the outside of the sensor, assembly or cable free
from damage?
‣
Perform a visual inspection.
Electrical connection Action
Are the mounted cables strain-relieved and not
twisted?
‣
Perform a visual inspection.
‣
Untwist the cables.
Is a sufficient length of the cable cores stripped, and
are the cores positioned in the terminal correctly?
‣
Perform a visual inspection.
‣
Pull gently to check they are seated correctly.
Are the power supply and signal lines connected
correctly?
‣
Use the transmitter wiring diagram.
Are all screw terminals tightened?
‣
Tighten the screw terminals.
Are all cable entries mounted, firmly tightened and
leak-tight?
‣
Perform a visual inspection.
In the case of lateral cable entries:
‣
Point cable loops downward so that water can drip
off.
Are all cable entries mounted on the side or pointing
downwards?