Turbimax CUS51D Technical data
Endress+Hauser 43
13 Technical data
13.1 Input
Measured variable • Turbidity
• Solids content
• Temperature
Measuring range
CUS51D-**C1 Application
Turbidity 0.000 to 4000 FNU
Display range up to 9999 FNU
Formazine
Solids content 0 to 5 g/l Kaolin
Filterable matter
Temperature –20 to 80 °C (–4 to 176 °F)
CUS51D-**D1 Application
Turbidity 0.000 to 4000 FNU
Display range up to 9999 FNU
Formazine
Solids content 0 to 300 g/l (0 to 2.5 lb/gal)
0 to 30 %
Solids content depending on the
selected application (see list)
Temperature –20 to 80 °C (–4 to 176 °F)
Measuring range with solids content:
For solids, the achievable ranges depend very much on the media that are actually
present and may differ from the recommended operating ranges. Extremely
inhomogeneous media may cause fluctuations in measured values, thus narrowing
the measuring range.
13.2 Energy supply
Power consumption 24V DC (-15 %/+ 20 %), 1.8 watt
13.3 Performance characteristics
Reference operating
conditions
20 °C (68 °F), 1013 hPa (15 psi)