Diagnostics and troubleshooting Turbimax CUS51D
38 Endress+Hauser
9 Diagnostics and troubleshooting
9.1 General troubleshooting
When troubleshooting, the entire measuring point must be taken into account:
Transmitter
Electrical connections and cables
Assembly
Sensor
The possible causes of error in the following table refer primarily to the sensor.
Problem Check Remedial action
Blank display, no sensor reaction Line voltage at transmitter?
Sensor connected correctly?
Buildup on optical windows?
Connect mains voltage.
Establish correct connection.
Clean sensor.
Display value too high or too low Buildup on optical windows?
Sensor calibrated?
Clean device.
Calibrate device.
Display value fluctuating greatly Is the mounting location correct?
Select a different mounting location.
Adjust measured value filter.
Pay attention to the troubleshooting information in the Operating Instructions for the
transmitter. Check the transmitter if necessary.