Operation Turbimax CUS51D
28 Endress+Hauser
8 Operation
8.1 Adapting the measuring device to the process
conditions
8.1.1 Applications
The sensor permits measurements in a wide variety of applications. The measuring
method is set automatically by selecting the relevant application.
Clear water application type
Application Method Measuring range
Formazine 135° - single-channel measurement 0 to 4000 FNU
Display range up to 9999 FNU
Kaolin 135° - single-channel measurement 0 to 5 g/l
TiO2 135°, 4-beam pulsed light 0.2 to 150 g/l
SiO2 135°, 4-beam pulsed light 5 to 100 g/l
0
10 20 30 40 50 60 100 120 140 160
0
1000
9999
2000
3000
4000
5000
6000
7000
8000
9000
70 80 90 110 130 150
FNU
D
C
g/l
B
A
A0050651
27 Clear water application type
A TiO2
B SiO2
C Kaolin
D Formazine
Solid application type
Application Method Measuring range
Thin sludge 135° turbidity, single-channel 0 to 5 g/l
Activated sludge 90°, 4-beam pulsed light 2 to 15 g/l
Excess sludge 135°, 4-beam pulsed light 3 to 50 g/l
Sludge, general 135°, single-channel (for low TS content) 0 to 50 g/l
135°, 4-beam pulsed light (for high TS content)
Digested sludge 135° turbidity, single-channel 5 to 100 g/l / 300 g/l