Turbimax CUS71D Commissioning
Endress+Hauser 23
Alarm delay echo loss establishes the amount of time
that the sensor must experience a loss of echo before
initiating the echo loss action.
You can enter a time from 0 to 255 minutes.
Reboot track performs a new initialization of the
sensor. The sensor starts in automatic mode and seeks
the interface.
It takes 3 to 5 minutes for the first measuring value.
If Sensor change is set to "on", the measured value at
the current output is set to hold. In this way you avoid
an error being reported at the process control system if
the sensor is replaced on site.
• Factory default measurement processing
Reset transmitter parameters to factory default
settings.
• Factory default sensor
Reset sensor parameters to factory default settings.