SPECIFICATIONS
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
13
SPECIFICATIONS
Specification Information
Transportation and
storage conditions
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
Relative humidity range of 10% to 95%, non-condensing.
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
61010-1 and CAN/CSA C22.2 No. 61010-1.
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
EN61326-1; EN61000-3-2; EN61000-3-3
IP-XO ordinary
Model EC1350BL EC1850BL
Intended use
Bottom cavity—warming blankets
TriMode cavity—warming blankets, irrigation or injection fluids
Weight
Net: 336 lb (152 kg)
Ship: 407 lb (185 kg)
Net: 457 lb (207 kg)
Ship: 528 lb (239 kg)
Storage cavity
capacity
TriMode cavity: 24 1-liter
bottles or 20 1-liter bags, or
9–12 blankets (3.9 ft
3
)
Blanket cavity: 24–30 blankets
(9.4 ft
3
)
TriMode cavity: 30 1-liter
bottles or 24 1-liter bags, or
10–14 blankets (5.4 ft
3
)
Blanket cavity: 30–40 blankets
(13 ft
3
)
Temperature
range
Irrigation fluids (IRR):
90°F to 150°F (32°C to 66°C)
Injection fluids (INJ):
90°F to 104°F (32°C to 40°C)
Blanket: 90°F to 160°F (32°C to 71°C)
Clearance
requirements
2” (102mm) from rear
1” (25mm) from top
1” (51mm) from sides
NOTE:
See topic Dimension
Drawings for EC1350BL Built-
In requirements.
4” (102mm) from rear
1” (25mm) from top
2” (51mm) from sides
1/2” (13mm) from bottom
E471516
Protective Earth
Ground Symbol