EasyManua.ls Logo

Enthermics EC1850BL - MAINTENANCE; Maintenance Schedule Overview

Enthermics EC1850BL
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAINTENANCE
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
39
MAINTENANCE
Maintenance Schedule
Daily For daily maintenance, do the following.
Check:
the air vents in the airflow insert panels are not obstructed (if applicable).
all fan guards are clear and not obstructed (if applicable).
the number of bottles/bags, or blankets as applicable do not exceed the
maximum capacity per shelf or basket.
Clean:
any spills with a clean, lint free cloth. See topic How to Clean the Warmer for the list
of approved cleaners.
Monthly For monthly maintenance, do the following.
Check:
the door gasket for tears and holes. Make sure that it is firmly attached to the
door. Check the seal when the door is closed.
the guards around the air temperature sensor are in place and fully secure to the
warmer.
the hardware securing the warmer(s) to the wall, if applicable.
Clean:
the outside of the warmer with stainless steel cleaner. See topic How to Clean the
Warmer for the list of approved cleaners.
vacuum (if applicable):
fan openings
fan sail switch
vent openings
Continued on next page
NOTE: Do not remove exterior panels from the warmer or attempt
repairs. The warmer has no user-serviceable internal components.
Only perform routine cleaning and maintenance procedures
specifically described in this manual.

Other manuals for Enthermics EC1850BL

Related product manuals