TROUBLESHOOTING
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
45
TROUBLESHOOTING
What to do if a Power Interruption Occurs
Background You may need to reset the warmer in the case that a power interruption occurs.
When the power is restored, the ON/OFF status indicator LED decimal
flashes .
The controller stores all settings and continues operation using these settings in
the event of a power interruption.
Procedure
To continue operation of the warmer, do the following.
Step Action
1. Press the standby button . The status indicator LED decimal goes out.
If an error code displays, refer to topic Error Codes.
2. Make sure the fluids and blankets are at the correct temperature before
using.
Result Resume operation of the warmer.
LED
NOTE: Inspect the fluids for their expiration date before
use.