INSTALLATION
Combin ation Blan ket/Flui d W armer Operat or s Manua l MN -399 40 Rev 3 8/20
23
How to Install the Warmer
Before you begin Make sure you have:
An appropriate lifting device and enough personnel to safely move and position
the weight of the warmer.
Requirements
The warmer must be installed on a level surface.
The warmer must not be installed in any area where it may be affected by
dripping water, high temperature, or any other severely adverse conditions.
Voltages
Continued on next page
WARNING: Tipping hazard.
Mount the EC1350BL to the floor before use.
Model V Ph Hz kW
EC1350BL 120
220
230
1
1
1
50/60
50/60
50/60
1.2
1.1
1.1
EC1850BL 120
220
230
1
1
1
50/60
50/60
50/60
1.3
1.2
1.3